TCDR-3A-1000 Thermal Conductivity Tester (Water Flow Plate Method)
PIC-(TCDR-3A-1000 Thermal Conductivity Tester (Water Flow Plate Method))
This instrument is suitable for testing the thermal conductivity of materials such as refractory insulation, ceramic fibers, felt, textiles, boards, bricks, etc. at different temperatures. Adding a frame can measure powdery materials. Widely used in material analysis and testing in universities, research institutions, quality inspection departments, and production plants.
The reference standard for this instrument is ASTM C518-04, Test Method for Determining Steady State Heat Flux and Heat Transfer Characteristics by Heat Flow Meter Method; ASTM C182-88 (2004) Standard Test Method for Thermal Conductivity of Insulated Refractory Bricks; ASTM C202-93 (2004) Standard Test Method for Thermal Conductivity of Refractory Bricks; ASTM C417-05 Standard test method for thermal conductivity of amorphous refractory materials; ASTM C201-93 (2004) Standard Test Method for Thermal Conductivity of Refractory Materials; GB10295-88 Determination of steady-state thermal resistance and related characteristics of insulation materials (heat flux meter method); YB/T 4130-2005 Test method for thermal conductivity of refractory materials (water flow plate method); GB/T 17911-2018 Test Method for Thermal Conductivity of Refractory Ceramic Fiber Products, etc.
Main technical parameters
Thermal conductivity range 0.0010-3W/mk
Test accuracy ± 5%
Hot surface temperature range 100-1000 ℃
Semiconductor refrigeration with calorimetry and temperature control method, optional with constant temperature water tank cooling, additional cost to be calculated
Thermal protection and temperature control method semiconductor refrigeration
Calorimetric plate size 100 x 100 (mm)
Heat flow meter size 100 x 100 (mm)
Sample temperature measurement points 4
Sample atmosphere, ambient air
Sample size 230 x 230 to 250 x 250 (mm); Thickness 10-100 (mm)
Measurement control method: Computer fully automatic testing, data processing, and the ability to generate detection reports for printing and output
Host power supply 1200 ℃ specification: voltage 220V; Frequency 50Hz; Total power ≤ 5KW
The multi-layer testing function instrument can perform multi-layer testing, and an additional testing thermocouple needs to be added for each additional layer (the cost of thermocouples outside of the standard configuration needs to be calculated separately).
The instrument configuration includes one TCDR-3A-1000 testing host, one set of software, one set of communication interfaces and data cables, one set of testing thermocouples (for three-layer testing), one roll of rubber water pipe, and the purchaser shall provide their own computer. If a constant temperature water tank is required, the cost will be charged separately